HFP03 heat flux sensor

HFP03 is an ultra sensitive sensor for measurement of small heat fluxes
HFP03 heat flux sensor for soil, walls and building envelope measurements
Ultra sensitive heat flux plate

HFP03 is an ultra sensitive sensor for measurement of small heat fluxes in the soil as well as through walls and building envelopes. The total thermal resistance is kept small by using a ceramics-plastic composite body. See also model HFP01: putting several HFP01 sensors electrically in series is an alternative for using HFP03.

Specifications
Measurand
  • heat flux
Measurement range
  • -2000 to +2000 W/m²
Sensitivity (nominal)
  • 500 x 10⁻⁶ V/(W/m²)
Would you like a personalised quote?

or contact us: info@huksefluxindia.com

Specifications

Measurand
  • heat flux
Measurement range
  • -2000 to +2000 W/m²
Sensitivity (nominal)
  • 500 x 10⁻⁶ V/(W/m²)
Sensing area
  • 64 x 10⁻⁴ m²
Sensor thermal resistance
  • 71 x 10⁻⁴ K/(W/m²)
Sensor resistance range
  • 10 to 32 Ω
Sensor thickness
  • 5.4 x 10⁻³ m
Uncertainty of calibration
  • ± 6 % (k= 2)
Rated operating temperature range
  • -30 to +70 °C
Cable diameter
  • 4 x 10⁻³ m
IP protection class
  • IP67
Standard cable length
  • 5 m
Options
  • longer cable length (10, 15, 20, 30, 40 m)
Options
  • longer cable, in multiples of 5 m, cable lengths above 20 m in multiples of 10
Manuals & Downloads

Description

HFP03 is an ultra sensitive sensor for measurement of small heat fluxes

Ultra sensitive sensor for measurement of small heat fluxes

HFP03 measures heat flux through the object in which it is incorporated or on which it is mounted, in W/m². HFP03 is specifically suitable for measurement of small flux levels, in the order of less than 1 W/m², for instance in geothermal applications. The sensor in HFP03 is a thermopile. This thermopile measures the temperature difference across the ceramics-plastic composite body of HFP03. A thermopile is a passive sensor; it does not require power. Using HFP03 is easy. It can be connected directly to commonly used data logging systems. The heat flux in W/m² is calculated by dividing the HFP03 output, a small voltage, by the sensitivity. The sensitivity is provided with HFP03 on its calibration certificate. 

Suggested use

  • extremely small heat fluxes
  • geothermal heat flux
HFP03 heat flux sensor
HFP03 is an ultra sensitive sensor for measurement of small heat fluxes
  • HFP03 is specifically suitable for measurement of small flux levels
  • the total thermal resistance is kept small by using a ceramics-plastic composite body
Would you like a personalised quote?

or contact us: info@huksefluxindia.com