Solar radiation calibration scale change
WRR solar irradiance scale is expected to go down
WRR contains a systematic error relative to SI. Therefore, the scale will be adapted. It will likely shift by 0.34 %. The change is relevant to climatology, solar resource assessment and solar system performance assessment.
Introduction
The WRR (World Radiometric Reference) has been used since 1980 as the scale representing the physical units of solar irradiance.
PMOD: designated institute under METAS
The WRR scale is maintained by PMOD. The formal status of PMOD in international metrology is “designated institute”. For details see the website of the Swiss Institute of Metrology METAS. In other words, PMOD is responsible maintenance of the highest standard for solar irradiance measurement with pyranometers and pyrheliometers.
What will happen to WRR
It has become clear that WRR contains a systematic error relative to SI. The WRR scale will be adapted to the new latest insights. It will likely shift by 0.34 %.
The WRR scale was too high, which means that all irradiance measured with traceability to WRR must be corrected to lower values in [W/m2]. With the correction from WRR to the new scale, the sensitivities of instruments in [V/(W/m2)] will go up by 0.34 %.
The name of the new scale is not yet known.
WMO does not recommend to correct past data. It does however recommend to store information with data obtained with pyranometers and pyrheliometers about the scale that is used. The decision on introduction of the new scale will be made at the WMO congress in 2027.
Literature reference for scale change
PMOD pyranometer calibration* | |
---|---|
Calibrated quantity | responsivity to solar irradiance |
Calibrated instrument | pyranometer |
Measurand | V/(W/m2) |
Uncertainty | 0.66 % |
Calibration method | pyranometer comparison to pyrheliometer |
Calibration conditions |
|
Calibration reference | World Radiometric Reference (WRR) |
Type of source | sun and clear sky |
* Reference: BIPM website |
What we do at Hukseflux to prepare
Hukseflux in its latest industrial pyranometers allows to add metadata, to indicate if calibration is traceable to WRR or to the new scale, to on-board calibration records and to on-board records of calibration history. All these records are stored in the instrument.
Download the PDF en read about solar radiation calibration scale change.